Answer: Dysfunctional Pump

The pump on this CTD was not running during the profile, causing noisy data during the upcast.

The figure above shows a zoomed-in view of 200 – 240 meters depth. Ship heave is shown in the oscillations in descent rate. If you look closely, a “bump” in salinity and temperature corresponds almost perfectly with each motion of the ship. This is a clear indication that the pump isn’t working – when a CTD has a functional pump, the temperature and conductivity sensors have a consistent flow of water regardless of descent rate and sea conditions. The resulting pumped data are smooth, with minimal effect from ship heave. Without a pump, the flow of water is dependent on the motion of the CTD package, where a faster descent rate will force more water through the plumbing. Data from the downcast are usually less affected, as the downward motion of the CTD package causes some flow through the plumbing. On the upcast however, water easily becomes entrained in the CTD’s plumbing, and may be unchanging, or will have a strong correlation with ship heave.

Read our Troubleshooting Guide for profiling CTDs to learn more troubleshooting techniques.